Analytical HPLC-A7840 system with 0.001 to 9.999 ml / min flow rate range, offers three method measurement: dual wavelength detection, wavelength time program and stop pump scanning. Integrated with a single and dual wavelength UV-Vis detector with a range of 190 to 700 nm to analyse sample’s absorption at different wavelengths and high-pressure pump achieves accurate solvent delivery with high-precision and low-fluctuation.
Available Range :
Detector Specifications :| Model No | HPD-A20 | HPD-A21 |
| Wavelength range | Single wavelength ( 190 to 800 nm ) | Dual wavelength ( 190 to 800 nm ) |
| Wavelength accuracy | ± 0.5 nm | ± 2 nm |
| Wavelength repeatability | 0.15 nm | 0.3 nm |
| Band width | 8 nm | 8 nm |
| Optical path | 10 nm | 10 nm |
| Deuterium lamp | 190 to 400 nm | 190 to 400 nm |
| Tungsten lamp | 400 to 800 nm | 400 to 800 nm |
| Baseline noise ( Static ) Au | ≤ ± 0.35 × 10-5 | ≤ ± 0.5 × 10-5 |
| Baseline drift ( Static ) Au / h | ≤ 1 × 10-4 | ≤ 1.5 × 10-4 |
| Minimum detection limit | ≤ 4 × 10-8 | ≤ 4 × 10-8 |
| Power | 100 W | 100 W |
| Power supply | 90 V / 50 to 60 Hz | 90 V / 50 to 60 Hz |
| Dimension | 460 × 315 × 180 mm | 460 × 315 × 180 mm |
| Weight | 9.5 kg | 9.5 kg |
| Pump Specifications |
| ||||||||||||||||||||||||||||||||
| Detector Specifications |
|
| Accessories no. | Accessories name | Specification | Unit |
| 1 | Analytical pump | 10 ml/min, 42 MPa | 2 |
| 2 | Analytical UV detector | Analytical flow cell included | 1 |
| 3 | Chromatography software | FDA compliance | 1 |
| 4 | Start-up kits | Accessories and tools | 1 |
| 5 | Static mixer | Analytical mixing | 1 |
| 6 | Manual injection valve | ID 4.6 × 250 mm 5 µm C18 kromasil | 1 |
Labtron Standard HPLC analyze complex mixtures efficiently and provides valuable data for both qualitative and quantitative analysis. They achieve high resolution with fine column packing and optimized flow rates. Their high-pressure pump offers precise mixing, sleek design and easy operation. The UV detector includes dual-beam monochromator, light battery, and deuterium lamp for accuracy. Our Standard HPLC is essential for precise compound analysis in pharmaceuticals, food, environmental studies, biotechnology, and forensics.
Labtron Analytical HPLC measures compounds in mixtures using column and detector system. They have deuterium and tungsten lamps that work together for both flexible and accurate detection. Measurement can be performed with three different methods for precise analysis. The system offers control via panel or computer for easy parameter and pump management. Our Analytical HPLC ideal for precise separation, identification, and quantification of compounds in various samples.