Generate Catalog For :

Analytical HPLC System LAHS-A10

labtron equipment ltd

Personalized PDF Catalog

Catalog Generated On : 09/11/2025


Description :

Analytical HPLC System LAHS-A10 features Quaternary Low Pressure Pump with 0.001-10 ml/min pump flow rate. Incorporated UV-Vis detector with a range of 190 to 800 nm that examines sample absorption at different wavelengths and a high pressure pump providing continuous stable flow. This HPLC system offers high speed accuracy, resolving power, detection, quantification, convenience and applicability to new sample types. 

Specifications :

Pump specification
TypeFlow Rate RangeFlow AccuracyFlow PrecisionMaximum PressurePressure accuracyPressure pulsationComposition rangComposition precision
Quaternary Low Pressure Pump0.001 to 10.000 ml/min, in 0.001ml increments≤ ± 0.2% (1.00 ml/min, H2 O, R.T., 8.5 MPa)RSD ≤ 0.06% (1.00 ml/min, H2 O, R.T., 8.5MPa)63 MPa≤ ± 3.0 %≤ 0.1 MPa0 to 100 %≤ 0.3 % RSD
UV/VIS Detector
Wavelength RangeWavelength RepeatabilityWavelength AccuracyBand WidthResponse TimeBaseline NoiseDriftPath lengthLight Source
190 to 800 nm≤ 0.1 nm≤ ± 1.0 nm6 nm0.0 s to 5 s≤ ± 0.25 × 10-5 AU(empty cell, 254 nm, 1.0s)≤ 0.5 × 10-4 AU/h (empty cell, 254 nm, 60 min)10 mm(analytical cell)Deuterium Lamp and Tungsten Lamp
Data-processing Workstation System contain the tables for regular samples
Equipped with Internal and external standard calculation process
Automatic display, print, export of data and starting of another program on completion on measurement
Observe with the condition of 21 CFR part 11 directive of the FDA
System Kits 1 Set
Combined A/D converters
Combined vacuum degasser and 4 channel
Power Supply AC 110 /220 V ±1 0 %, 50/60 Hz
Dimension 710 × 583× 500mm; 770 × 450 × 700 mm
Weight 156 kg

Features :

  • UV-Vis detector with a range of 190 to 800 nm
  • Consist of Quaternary Low Pressure Pump
  • Advance and intellectual HPLC system
  • Excellent Flow Accuracy and Pressure Accuracy
  • Column oven with greater capacity of 3 column with maximum thickness ID10 × 250 mm
  • Fast and stable UDP communication protocol, can be used for instrument remote control
  • Enhancement of up to 5 mm stroke of plunger rod with good performance and high pressure constant flow pump or parallel pump head
  • Complies with FDA 21 CFR part 11

Applications :

High Performance Liquid Chromatography system has been widely used in Research and Development, pharmaceutical, food nutrient analysis of amino acids and sugars, detection of food contaminants, environmental pollutant and chemical product purity.


labtron equipment ltd

www.labtron.co

Labtron Equipment Ltd. Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.co | Website: https://www.labtron.co/ | Phone: +44 2080 043 608