Generate Catalog For :

Automatic Cleveland Open Cup Flash Point Tester LOFP-A12

labtron equipment ltd

Personalized PDF Catalog

Catalog Generated On : 19/11/2025


Description :

Our LOFP-A12 Flash point tester is equipped with English man-machine interface. It can preset expected flash point temperature and other parameters which is reasonably designed, safe and convenient to operate. It is designed and made as per ISO 2592, ASTM 92 “Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods).

Specifications :

Ambient Temperature 10 ℃ ~ 40 °C
Relative Humidity ≤ 80 %
Temperature Measurement- Full Scale Ambient temperature to 300 °C
Temperature Measurement- Repeatability ≤ 8 °C
Temperature Measurement- Reproducibility ≤ 16 °C
Temperature Measurement- Resolution 0.1 °C
Test accuracy 0.002
Precision 0.005
Flame 3 to 4 mm
Dimensions 550 x 470 x 410 mm
Power Consumption ≤ 500 W
Power Supply AC 220 V ± 10 %, 50 Hz

Features :

  • Automatic program temperature controlling
  • Automatic meteorological pressure standardization and altered value correction
  • Electric flame ignition mode
  • LCD display
  • It can print out test data automatically
  • Convenient and rapid

Applications :

Used to determine flash point of petroleum products and bituminous materials, excepting fuel oil and materials having a flash point of lower than 79 °C


labtron equipment ltd

www.labtron.co

Labtron Equipment Ltd. Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.co | Website: https://www.labtron.co/ | Phone: +44 2080 043 608