Atomic Absorption Spectrophotometer LAAS-A13


Share :


Atomic Absorption Spectrophotometer LAAS-A13


Flame Atomic Absorption Spectrophotometer LAAS-A13 is a microprocessor controlled instrument based on analysis of metal ions by flame and graphite furnace atomizer with their absorption at wavelength of 190 to 900 nm. Czerny-Turner monochromator and variance of bandwidth provides user flexibility. It is provided with deuterium hollow cathode lamp to exclude interference of molecular absorption and Titanium Burner to increase the analysis efficiency and precision of flame. Its double beam radiation sources enables good basic line stability. Graphic analysis system provides data on analysis of entire process e.g. measured value, temperature, procedures, time, etc. It has great applications in research, medicine, environmental science and other industries.



Available Range :

Wavelength range : 190 - 900 nm Wavelength range : 180 - 900 nm Wavelength range : 180 – 900 nm Wavelength range : 180 – 900 nm Wavelength range : 190 nm ~ 900 nm Working spectral range : 190 to 900 nm
Working spectral range190 to 900 nm
Spectral Bandwith0.1, 0.2, 0.4, 1.0 and 2.0 nm
Wavelength accuracy± 0.15 nm
Wavelength repeatability≤ 0.04 nm
Baseline stability≤ 0.002 Abs / 30 min
Characteristic concentration of copper0.02 μg / ml / 1 %
Detection limit of copper0.004 μg / ml
Measurement Repeatability0.005
Grating1800 lines / mm
InflamerAll-metal titanium burner
AtomizerEfficient glass atomizer
Lamp stand8 lamps automatic turret, automatic alignment
D2 Background calibration ability1 A ≤ 30 times correction ability
Dimensions700 × 550 × 450 mm
Net weight75 Kg
Power supply220 V ± 22 V AC
Acetylene air burner100 mm
Ignition dynamic baseline drift≤ 0.006 Abs / 30 min
Characteristic viscocity≤ 0.025 μg / ml / 1 %
Standard deviation of the accuracy≤ 0.5 % (Cu, absorbance > 0.8 A) (detection limit Cu ≤ 0.008 μg / ml)
Protection systemCan inevitably cut off gas if pressure is low, power cut-off, flame out and alteration of the burner
  • Integrated flame/graphite furnace atomization system, changeable with flame emission burner
  • Turret for 8 pre-aligned Hollow Cathode Lamps fitted with pre-heating, and individual pre-centred lamps
  • PID and Dual curve mode light-controlled temperature control technique ensures fast heating and analytical sensitivity
  • Fully automated wavelength scanning and peak searching
  • Advance software for complete parameter settings
  • Reliable fully automatic graphite furnace analysis
  • Alarm feature to protect fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
  • Analysis of elements by flame absorption method and emission method
  • Strong database system which possesses more than 500 data self-storage
  • Results printed as parameter, data result and diagram
It is suitable to analyse trace amount of element from mixture, also in forensics and chemistry, biological science to detect toxic elements from blood and drug samples.
Accessories no. Name
1 PC workstation
2 Printer
3 Oil free Air compressor
4 Acetylene reducing valve
5 Cu - Hollow cathode lamp
6 Air filter
GRAPHITE FURNACE
Maximum temperature 3000 ℃
Temperature increasing speed ≥ 2000 ℃ / S
Characteristic quantity (Cd) ≤ 0.5 × 10 to 12g , (Cu) ≤ 0.5 × 10 to 11g
Accuracy (Cu) ≤3 % , (Cd) ≤ 3 %
Dimension 550 × 450 × 300 mm
Net weight 65 Kg
Safety system Protection for Overload current, reduced air pressure, low cooling water flow
Power supply 220 V ± 22 V AC, 7000 W